Total volume (liter)
|
Working volume (liter)
|
Minimum working volume (liter)
|
Aspect ratio total volume
|
Aspect ratio working volume
|
|
1 liter
|
1
|
0.9
|
0.3
|
1.9
|
1.5
|
|
2 liter single wall
2 liter jacketed
|
2
2
|
1.7
1.7
|
0.5
0.5
|
2.2
|
1.9
1.9
|
|
3 liter single wall
3 liter jacketed
|
3
3
|
2.7
2.7
|
0.5
0.5
|
1.7
|
1.5
1.5
|
|
5 liter single wall
5 liter jacketed
|
5
5
|
3.4
3.4
|
2.0
2.0
|
1.6
|
1.1
1.1
|
|
7 liter single wall
7 liter jacketed
|
7
7
|
5.4
5.4
|
2.0
2.0
|
2.3
|
1.8
1.8
|
|
15 liter single wall
15 liter jacketed
|
15
15
|
12
12
|
2.5
2.5
|
1.9
|
1.5
1.5
|
|
20 liter single wall
|
20
|
16
|
2.5
|
2.5
|
2.0
|
|
Drive system
|
Direct drive, lipsealed or magnetically coupled
|
|
Maximum stirrerspeed (rpm)
|
Standard range is 50 - 1250.
1, 2 and 3 liter systems can be supplied with 2000 rpm motor
|
|
Impellers
|
Rushton and marine with outside diameters 45mm, 60 mm 75 mm or 85 mm
|
|
Gas sparger
|
Porous sparger or L-type sparger
|
|
Gas overlay
|
Yes
|
|
Exhaust gas
|
Water cooled exhaust gas condenser
|
|
Sampling
|
Fixed height or height adjustable sample pipe with optional sampling system
Sample pipe internal diameters choices are: 1.7 mm, 4 mm, 6 mm or 10 mm
|
|
Draining
|
Drain pipe
|
|
Additions
|
Triple or single inlet ports and optional micro liquid injectors
|
|
pH
|
Measurement: 12 mm classic pH sensor
Control: via acid pump or CO2 gas (rotameter or MFC) in combination with alkali pump
|
|
DO2
|
Measurement: 12 mm classic polarographic DO2 sensor
Control: via a combination of N2, Air, O2 (Rotameter or MFC) and agitation or nutrient addition pump
|
|
Temperature
|
Measurement: Pt-100 sensor in thermowell in topplate
Control: cooling and/or heating jacket via bioreactor wall or via internal heat exchanger
|
|
Foam
|
Measurement: Height adjustable conductivity based foam sensor
Control: Anti foam addition pump
|
|
Level
|
Measurement: Height adjustable conductivity based level sensor
Control: pump for liquid addition or removal
|
|
Optional inlets
|
Septum, chemostat tube, liquid entry system
|